DEVELOPMENT OF A THERMAL CONDUCTIVITY MEASURlNG APPARATUS FOR NON-METALLIC SOLIDS

Show simple item record

dc.contributor.author ANJORIN, S. AYODEJI
dc.date.accessioned 2022-12-13T12:44:18Z
dc.date.available 2022-12-13T12:44:18Z
dc.date.issued 1996-07
dc.identifier.uri http://196.220.128.81:8080/xmlui/handle/123456789/5541
dc.description M.ENG THESIS en_US
dc.description.abstract A thermal conductivity measuring apparatus for non-metallic solids (poor conductors) based on modified Lee's disc method has been designed, constructed and tested. Specimens of concrete, glass, portland cement, asbestos and corrugated cardboard were prepared in the form of discs each of diameter 50mm but of varying thicknesses. The average thermal conductivity values obtained for these specimens over a heat source temperature range of 25°C - 60°C were found to be 1.285, 0.737, 0.327, 0.121 and 0.072 Wm-1 K-1 respectively. The values of thermal conductivity obtained for the various specimens were compared with those given in the literature for the purpose of determining the effectiveness of the apparatus. The differences in values ranged between 1.55 - 11.03 per cent thus giving credence to the effectiveness of the apparatus. en_US
dc.description.sponsorship FEDERAL UNIVERSITY OF TECHNOLOGY AKURE en_US
dc.language.iso en en_US
dc.publisher FEDERAL UNIVERSITY OF TECHNOLOGY AKURE en_US
dc.subject NON-METALLIC SOLIDS en_US
dc.subject THERMAL CONDUCTIVITY en_US
dc.title DEVELOPMENT OF A THERMAL CONDUCTIVITY MEASURlNG APPARATUS FOR NON-METALLIC SOLIDS en_US
dc.type Thesis en_US